55
Instructions for use for accessories
The device has two dierent sta call operating modes:
Switched off
Static
without
offalarm *)
Dyn.
without
oalarm *)
Alarm
Opera-
tion
Alarm
Opera-
tion
Switched offSwitched on
Operating alarm
1 sec.
*
In “static without o alarm” mode, the sta call can be disabled by pressing the OK key.
Technical data
Mode: Status
red LED
Status Changeover Contact
(wire color white - green)
Status Changeover Contact
(wire color white - brown]
Operating: o closed opened
Alarm: on opened closed
Accessory Parameter Value
Interface lead 12 V CP
(8718020)
Dimensions in mm
(W x H x D)
Cable length in mm
Approx. 75 mm x 43 mm x 15 mm
Approx. 2568 mm
Interface lead sta callCP
(8718030)
Dimensions in mm
(W x H x D)
Cable length in mm
Approx. 75 mm x 43 mm x 15 mm
Approx. 2553 mm