Loading...
Instructions for Use
Code: IU.01.09.0007
Version / Revision: 6.001
Page: 87 de 143
Ultrasonic washer
Model: E0201
Weekly:
Clean the internal surfaces of the tank (more details below);
Monthly:
Check the state of the slap joints.
Details:
Cleaning the bath surface:
If the container contains inlaid material, rub the FACTS® descaler (Baumer Code HER-501) on the surface. The tank
temperature must be low enough to allow contact with the operator's hands:
Wait 15 to 20 minutes and pack the container with a non-metallic brush (ideal brush with nylon
bristles). Remove the product with a flush and / or wet cloth;
Be careful not to leave fur on the surface;
The frequent use of FACTS® (Baumer Code HER-501) is recommended for surfaces with excess of
encrusted material.
Cleaning of drainage drain (Figure 11):
Pull it upwards;
Clean the grid;
Position the grid again in the drain.
Figure 7 Grid
Cleaning of exterior finishes:
Carry out cleaning with a soft cloth or sponge (non-abrasive), wet in hot or warm water and with mild
soap;
In case of doubt about the cleaning materials that will be used in the Ultrasonic washer, consult the
supplier of the same about the application in stainless steel.
When cleaning always use the appropriate PPE.
Make sure the ultrasonic washer is off.
Never use abrasive sponges, steel wool or metal instruments that could damage the surface.
The cleaning of the external trim and the interior of the chamber should be done using a soft cloth and
non-aggressive solutions to the stainless steel.
Some surfaces may be hot. Before starting cleaning, wait for the temperature drop to safe values (less
than 40°C).