07-19-B1-688 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP SERVICE MANUAL 5-5
Configuration Option Feature Modifying the Configuration Option Settings
17. Programmed Delivery Profile: Determines the memory to store the
programmed delivery profile.
Note: Ramp PDP option is available only on pumps running software
versions 1.09 or later.
1: Disabled
2: 5 hour memory
3: Semi-permanent memory
4: Permanent memory
5: Ramp PDP (Profile is
saved in permanent memory.
When a new PDP is entered,
the pump reverts to the PDP
in permanent memory once
the cycle has been
completed.)
Disabled
18. Time Setting: Set the real time clock in hours and minutes. (military
time 00:00 - 23:59)
—— Central
Standard
Time-CST
19. Date Setting: Set the date using the Month/Day/Year format. —— Current
date-CST
Table 5-1 Configurable Settings — continued
Parameter Description Setting Options
Factory
Settings