07-19-B1-688 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP SERVICE MANUAL 6-1
Chapter 6
Chapter 6 Routine Maintenance
Overview
This chapter contains a table describing preventive maintenance which should
be performed on the FLO-GARD 6201 Single Channel Volumetric Infusion
Pump. The maintenance procedures outlined in this section may be performed
in the hospital. When an abnormal condition occurs which is not correctable
by performing the following procedures, remove the pump from service and
troubleshoot it in accordance with Chapter 7, “Troubleshooting”, or return it
to Baxter for repair.
Cleaning
The pump should be cleaned as soon as possible after each use to minimize
the accumulation and hardening of spilled solutions. The case and front panel
may be cleaned with a soft cloth or cotton swabs dampened with a properly
diluted cleaning agent listed in Table 6-1.
Be sure to follow the manufacturer’s dilution instructions for concentrated
cleaners where applicable. Do not spray cleaning agents directly onto the
inside of the pump door, the pump mechanism, and the front panel film. When
these areas require cleaning, wipe carefully with a soft cloth, sparingly
dampened with a cleaning agent listed in Table 6-1. When solution spillage
onto the pumping mechanism or front panel occurs, it must be cleaned
immediately. If necessary, contact Product Service at 1-(800)-THE-PUMP.
! WARNING !
If the pump has been dropped or appears to be damaged, it should be taken out of
service and carefully inspected.
CAUTION
Do not submerge the pump into liquids of any kind.