07-19-B1-688 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP SERVICE MANUAL 8-19
Removal/Replacement Procedures Disassembly/Reassembly
Replacement of Safety/Slide Clamp Assembly
1. Remove pump door cover in accordance with “Replacement of the Pump
Door or Pump Door Assembly” on page 8-11.
2. Separate the front and rear housings in accordance with “Separation of
Front and Rear Housings” on page 8-5.
3. Remove the pump head in accordance with “Replacement of Pump Head
Assembly” on page 8-12.
4. Press the safety clamp pin (Figure 11-9 , item 6) to close the safety
clamp.
5. Remove the screw (Figure 11-7 , item 2) on the right side of the safety
clamp arm cover (Figure 11-7 , item 10) and remove the arm cover.
6. Close the pump door.
7. Place the pump head face down and remove the two screws
(Figure 11-9 , item 35) to remove the sensor assembly.
8. Remove the two screws (Figure 11-9 , item 30) on the bottom of the
safety/slide clamp assembly to release the motor holding plate
(Figure 11-9 , item 8).
9. Remove the six screws (Figure 11-6 , item 19) on the finger box and lift
the finger and motor assembly off of the base plate.
10. Remove the flexible cable at the back of the slide clamp assembly
(Figure 11-9 , item 29) from the connector on the terminal board. Be sure
to pull the reinforced part so as not to damage the flexible cable.
11. Remove four screws (Figure 11-8 , items 28 and 40) securing the safety/
slide clamp assembly and remove it from the pump head base plate.
12. Press the safety clamp pin (Figure 11-9 , item 6) on the replacement
assembly to close the safety clamp, with the safety clamp seal dangling,
install the replacement safety/slide clamp assembly into rear of the pump
head base plate and secure with four screws (Figure 11-8 , items 28 and
40). Make sure safety clamp seal (Figure 11-9 , item 25) is in place and
moves freely side to side with the safety clamp arm. Tightening torque
for the four screws is 6 kgf-cm (5.2 in-lb).
13. Inspect for a white or yellow dot on the side of the slide clamp assembly
(nearest the middle of the device) and on the FPC terminal PCB (see
Figure 8-3 for location). If a dot is present, proceed to step 14.