07-19-B1-688 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP SERVICE MANUAL 7-11
Troubleshooting Failure Identification Code Troubleshooting Table
Automatic Test Mode 8: Air Sensor Test Mode
Upon entering this mode, the PRI RATE is set to 30, PRI VTBI = 100, SEC
RATE = 600, and SEC VTBI = 100. Pressing the
TIME key selects the pump’s
air sensor sensitivity (MIN or NORM) without using the configuration option.
The air sensor selection is displayed in the message display when the pump is
started. All other keys function normally except for the
TIME key. Rate and
VTBI information for both primary and secondary mode may be changed.
Automatic Test Mode 9: Elapsed Time Test Mode
The elapsed pumping time is displayed in the first line of the message display
in this test mode. Momentarily pressing the
TIME key selects two
pre-programmed primary mode test conditions. The first is a PRI RATE of
125 mL/hr with a VTBI of 100 mL. The second is a PRI RATE of 499 mL/hr
with a VTBI of 100 mL. The elapsed time counts only when the pump is
running. Stopping the pump and then restarting it resets the timer to zero. All
other keys function normally except for the
TIME key. When the VTBI has
been delivered, the pump enters KVO mode.
Automatic Test Mode 0: Downstream Occlusion Test Mode
In this mode, pressing the TIME key allows you to select the pump’s
downstream occlusion level sensitivity without entering the configuration
mode. The selected occlusion sensor level (1, 2, or 3) is momentarily shown
in the second line of the message display until the test mode is exited. After
exiting test mode, the downstream occlusion sensor level reverts to the
original configuration setting. Rate and VTBI information may be
programmed manually.
Failure Identification Code Troubleshooting Table
Use Table 7-3 to determine the corrective action necessary to resolve failure
codes. The causes of each failure code are listed in the order in which they are
most likely to occur. Perform the corrective action items in the order in which
they are listed and retest the pump after each action. All replacement
procedures are contained in Chapter 8, “Removal/Replacement Procedures”.
There are two types of Failure Identification Codes: Alarm (codes 1-19) and
Failure (codes 20 and above). Alarm codes report malfunctions generally
correctable by the operator. Failure codes may require servicing of the pump.
A code of 0 indicates that less than 10 alarms/failures have occurred since the
log was cleared. The time and date are also 0.