EasyManua.ls Logo

Berthold Duo Xpert - Calibration Settings: Background

Berthold Duo Xpert
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Main Menu Device Setup LB 470 Level
102
56925BA2 Rev. 03, 11/2017
Calibration Settings: Background
Device settings | Setup | Calibration | Calibration Settings
The background count rate (Fig. 82, item 1) indicates the natural background radi-
ation of the detector, if no radiation source is installed. This count rate is
compensated for by the system. The best approach for your background measure-
ment is dependent on the situation on site and on the type of radiation source.
NOTICE
A closed shield also results in measurable residual radiation, which can falsify the
measurement of background radiation. For this reason, it is recommended that
the detector (Fig. 81, item 4) be determined at a suitable distance (approx. 10 m)
or behind a thick concrete wall (Fig. 81, item 2).
In the case of measuring arrangements with point sources, it is recommended to
place the shield with spotlights at a suitable distance (approx. 10 m), or behind a
thick concrete wall.
1
2
3
4
Radiation shield (closed)
Concrete wall
Distance
Rod detector
Fig.
81 Conditions during background determination

Table of Contents