EasyManua.ls Logo

Berthold Duo Xpert - Adjust: Standard Adjust

Berthold Duo Xpert
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LB 470 Level 7 Main Menu Device Setup
56925BA2 Rev. 03, 11/2017
111
Adjust: Standard Adjust
Device settings | Setup | Calibration | Adjust
Use the functionality after source exchange or after entering a theoretical, normal-
ized multi-point calibration (e.g. a Radical calculation) to adjust the table in the
measuring set.
That will keep the shape of the curve, because all calibration points are extrapo-
lated as percentage of the difference between the old- and the new calibration
points.
1
2
Adjusted curve (linear calibration type)
Adjusted curve (exponential calibration type)
Fig.
91 Adjusted curves
NOTICE
During the Standard Adjust, level and pressure must remain constant.
1
2
3
4
5
6
Button < Perform adjust now >
Input field
Read-In-Time
Input field
Process value
″ %
Button < Ok >
Button < Start >
Button < Cancel >
7
8
9
Input field
Count rate
[cps]
Process bar
Button < Stop >
Fig.
92 Standard adjustment

Table of Contents