EasyManua.ls Logo

Berthold Duo Xpert - GPC calibration: Factor M; Capturing the count rates

Berthold Duo Xpert
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Main Menu Device Setup LB 470 Level
122
56925BA2 Rev. 03, 11/2017
GPC calibration: Factor M
For complex applications, differences in geometry between level measurement and
density measurement may exist and/or highly different absorption coefficients (e.g.
by different nuclides) may occur. In this case, an adjustment for at least 2 (up to 10)
different gas densities is necessary to determine the manual correction factor (M).
No adjustment needs to be performed for standard applications.
1
2
3
4
5
6
List GPC count rates
Input field Factor M
Button <Calculate>
Button <+> (manual adding count
rates)
Button <Add Reference> (current
calibration value)
Button <Close>
7
8
9
10
11
12
Button < - > (removal of the point)
Button <Ok>
Input field level detector
Input field GPC detector
Measuring Duration
Button <Start>
Fig.
102 GPC calibration: Calculate factor M
Capturing the count rates
IMPORTANT
If it is not possible to capture the count rates, the values may be entered man-
ually into the input fields (Fig. 102, item 9, item 10).
1. Click on the < + > button (Fig. 102, item 4) to add a new point.
2. Click on the input field "Read-In-Time" (Fig. 102, item 11) and enter a meas-
urement time. Confirm with the Enter key.
3. Click on the button <Start> (Fig. 102, item 12). Make sure that the actual
level in the container does not change during the measurement time.
Repeat the capture for at least one other level.
4. A new point is added when clicking on the button <Add Reference> (Fig.
102, item 5), which takes over the count rates of the measurement during
calibration.
2
1
3
4
5
7
6
8
12
11
10
9

Table of Contents