7 Main Menu Device Setup LB 470 Level
56925BA2 Rev. 03, 11/2017
Signal processing: XIP (Radiation Interference)
Device settings | Setup | Signal Condition | XIP
This function allows you to take interference (XIP) into consideration. Measure-
ment jumps that influence the process can arise through interference. Only rapid
increases are considered.
If detection (Fig. 112, item 1) is activated, the last valid measured value is frozen.
This value determines the wait time for the
measured value generation. The change does
not affect the measurement above this time.
The valid measurement value is frozen at this
time after detection of interference.
The ″I
0
factor″ determines the recognition crite-
ria for interference.
The standard deviation is set in the field ″RI
sigma″ (radiation interference sigma).
″
″
measurement delay
Input field hold time
Input field I_O factor
Input field RI Sigma
112 Signal Condition (Radiation Interference)
Rapid Switch and Interference detection must not be activated simultaneously.