EasyManua.ls Logo

Berthold Duo Xpert - Detector Temperature Alarm Settings

Berthold Duo Xpert
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Main Menu Device Setup LB 470 Level
146
56925BA2 Rev. 03, 11/2017
Detector Temperature Alarm Settings
Device settings | Setup | Alarms | Det.-Temp. Alarm Settings
The values for the detector temperature (max. and min.) can be set in the submenu
PV Det.-Temp. Alarm Settings.
When there is exceeding or falling below the switching point, an event message
appears in the status display. If a digital output Detector temperature alarmis
assigned under Function, the relay switches.
1
2
3
4
Selection box “Min. Active”
Selection box “Max. Active”
Input box “Switch point Min.”
Input box “Hysteresis Min.”
5
6
7
8
Input box “Switching point Max.”
Input box “Hysteresis Max.”
Event message “Error”
Event message “Outside the specifi-
cation”
Fig.
129 Det.-Temp Alarm Settings
Setting max. temperature alarm:
1. Activate the check box (Fig. 129, item 1).
1. Click on the input field (Fig. 129, item 2) to enter a switching point.
2. Confirm with the Enter key
The value was changed.
Set min. temperature alarm:
1. Activate the check box (Fig. 129, item 3).
2. Click on the input field (Fig. 129, item 4) to enter a switching point.
3. Confirm with the Enter key
The value was changed.

Table of Contents