7 Main Menu Device Setup LB 470 Level
56925BA2 Rev. 03, 11/2017
An update of the firmware of the detector may take 1 hour and may only be
performed by qualified specialists.
The current software versions can be downloaded from the Berthold website
(www.berthold.com).
In order for the system to detect the update file it must not be located in an
index in the USB storage device.
Perform firmware update
1. Save the current update file of the firmware of the detector on a USB stor-
age device.
2. Connect a USB storage device to the device (Fig. 3, item 5).
3. The USB storage device is recognised by the system after a few seconds and
the <Firmware Update> (
Fig. 74, item 4) button can be clicked.
4. Click on the button <Firmware Update> (Fig. 74, item 4).
5. The update is performed.
Berthold recommends a test or a calibrating the current outputs whenever if a
software update has been carried out.