LB 470 Level 7 Main Menu Device Setup
56925BA2 Rev. 03, 11/2017
Detector settings: Detector Service | Reset Detector
Device settings | Setup | Sensors | [NAME DETECTOR] | Detector Service | Reset De-
tector
In the submenu ″Reset Detector″, the detector can be restarted and be reset to the
factory settings.
Button Restart
Button Factory settings
Window "Warning: Reboot"
Window "Warning: Factory Reset"
78 Reset Detector
The measurement is interrupted during a restart.
1. To restart the Detector, click the button <Reboot> (Fig. 78, item 1).
A window with a warning "Reboot" (Fig. 78, item 3) opens.
2. Click on <Yes> to confirm.
The device is restarted.
All custom configuration settings will be lost with a reset to factory settings!
1. To reset the detector to the factory settings, click the button <Factory reset>
(
Fig. 78, item 2).
A window with the warning Factory settings (Fig. 78, item 4) opens.
2. Click on <Yes> to confirm.
The device is reset to factory settings and restarts.