KB (E4 + E6) 02/2017 page 45/148
Performance after completing the program
The controller automatically changes to the “Fixed value” operation mode.
Before starting the program, check the temperature setpoint entered in the “Fixed value” operation mode.
After end of the program, the temperature will equilibrate to this value.
CAUTION
Too high or too low temperature after the program ends.
Damage to the charging material.
Before starting the program, check the temperature setpoint of “Fixed value” operation
and if necessary adapt it.
9.2 Cancelling a running time program
To cancel a running time program, go to Menu > Programs > Time program > Stop.
To cancel a running time program, you can also go to Quick menu > Time program > Stop.
The controller returns to the initial view.
Performance after manual program stop
The controller automatically changes to the “Fixed value” operation mode.
Before starting a program, check the temperature setpoint entered in the “Fixed value” operation mode.
After cancelling the program, the temperature will equilibrate to this value.
CAUTION
Too high or too low temperature after cancelling the program.
Damage to the charging material.
Before starting the program, check the temperature setpoint of “Fixed value” operation
and if necessary adapt it.