WARNING
S
The control panel screen will be damaged if cleaned with a
solvent containing solution.
Never use solutions containing Chlorine such as bleach on Stainless Steel.
Do not use alcohol or similar cleaning agents as it will adversely affect the
Hydrogen Peroxide Sensor
6 PREVENTATIVE & SCHEDULED MAINTENANCE
6.1 Operator Maintenance
6.1.1Cleaning
The QUBE exterior surfaces are manufactured from Polypropylene, ABS/PC,
painted Polyurethane, Epoxy coated Aluminium Alloy and Stainless Steel.
To keep the QUBE exterior surfaces clean the following procedure is
recommended:
1. Using a mild detergent solution (non abrasive) wipe the external surfaces with
a damp cloth taking care not to wet the control panel. All electrical
connections should be capped.
2. Rinse cloth and wipe thoroughly.
3. Dry surfaces with a lint-free clean cloth.
If cleaning near the mains power input then the QUBE should be switched off;
otherwise power can be left on.
Compatibility of different cleaning solutions with the QUBE materials of
construction cannot be guaranteed.
For internal cleaning Bioquell recommend use of KLERWIPE-CR (wipes), Biocides
A, B, C or D, from www.shieldmedicare.com.
Ensure all internal electrical connections are capped when cleaning the chamber
interior.