No. Module Description Condition *Action User Action
131 QHPV QSDM right-
hand door
closed
If in multi chamber cycle and door closed. 2 Open door, re-
run cycle
132 QHPV RH level not
achieved
If in cycle and RHAlert limit not achieved by
end of gassing.
1,2 Re-run cycle
133 QHPV QMTD left-hand
recirculation
fault
If QMTD present and in multi chamber cycle
and over pressure switch is made.
2 Contact
Bioquell
134 QHPV QMTD right-
hand
recirculation
fault
If QMTD present and in multi chamber cycle
and over pressure switch is made.
2 Contact
Bioquell
* Action: 1 - Aborts QHPV-only decontamination cycles
2 - Aborts full-system decontamination cycles
3 - Ends aseptic hold
Advice on what to do in the event of an alarm is available from Bioquell UK or
your local Bioquell agent.
Contact details: csts@bioquell.com; or Tel: +44(0)1264 835835.
When contacting Bioquell please quote the alarm number.
7.2 Troubleshooting
QUBE system
Check power lead is properly
connected.
Check power is switched ON at
mains.
Check QUBE power switch is ON.
Contact Bioquell Service Provider.
working
Check mains power indicator is
illuminated. Check front access
window is shut.
Contact Bioquell Service Provider to
replace light strip.
movable
Check castor locks have been
released and that stays are not down.
Check castors are not fouled by
debris or cables.
Check if castors seized or damaged.
Clear as necessary.
Contact Bioquell Service Provider.
not operate
Check correct paper has been used.
Check green light on printer is lit.
Check paper feed.
Replace paper roll if necessary. The
green light will flash.
Press paper advance key on printer.
Check paper was the correct side up.
An appropriately trained person may
check all control cable connections.
Contact Bioquell service provider.
password
enter password in accordance with
Manual instructions.
Re-set password by administrator.
rights.
extended
If the cycle time increases, check that
alcohols, IPA or other hydrocarbons
are not present in the atmosphere,
being used to clean the Qube or being
used to wipe items prior to being
placed in the Qube
level sensor to verify the end
of the cycle, see section 4.3.4.6
(The Hydrogen Peroxide sensor is
adversely affected by hydrocarbons,
resulting in sluggish response and a
zero offset of the ppm reading)