Note:
. The cycle
used
should be validated for the presentation and load
being
bio-decontaminated.
APPLICABLE TO USA ONLY
The
QUBE must be operated in accordance with
the Bioquell Hydrogen Peroxide sterilant labelling. It is Federal offence to use
pesticides in a manner inconsistent with their labelling.
4.3.2Loading a Hydrogen Peroxide Bottle
If attempting to start a gassing cycle and there is either no bottle loaded,
insufficient Hydrogen Peroxide within the bottle or it has expired, then the
button will not be visible.
Figure 29: Opening bottle module door
The bottle module door can only
be opened manually if there is no
bottle loaded. Otherwise, unlock
by selecting the bottle changing
function, , on the touch
screen.
Figure 30: Location of bottle register
Unscrew the cap and keep safe.
Failure to remove cap will damage
equipment. Then, carefully place the
H
2
O
2
bottle upright into the holder
ensuring the RFID symbol on the
bottle is at the back of the module then
the bottle should locate on the register
(as shown in Figure 24); slight rotation
may be necessary until it drops down.
This ensures that once installed the RFID
tag on the bottle is aligned correctly with
the reader within the QUBE unit.