1 691 606 210 2015-07-29| Robert Bosch GmbH
Fault procedure | BSA 43xx | 69 en
9.3 BNet Office and SDL 43xx
Error message Explanation/remedy
Print-out error. No valid order data available.
If this error message occurs with every print-out, end all programs. Switch the
system off at the master switch box and back on again. Re-start the PC pro
-
grams.
Host manager not available!
or
Data service not available!
Single-user workstation:
Check whether the "Host manager" and "Data service" program modules are
displayed in the order bar at the bottom right.
If not, end all programs and re-start the software modules. The programs
should then be shown as a symbol in the order bar.
If not, Data service was probably not selected on installation, so that
configurations and order editing were not yet possible.
Inform your Service engineer.
Network:
The PC or server on which Data service is installed in the network is not run
-
ning or has not been logged onto your PC.
Driver initialization unsuccessful. The COM port of the serial interface cannot be opened, as it is being used by
another application.
No communication established with field bus users. Check whether the entire system is running. Is the
green LED on the BNet converter box flashing?
Check whether all BNet connectors are firmly plugged in.
Was the system inadvertently started after the PC soft
-
ware? The PC software must be started last.
Check whether the analog display is outputting an error or is
still busy. Most messages can be acknowledged with
STOP or
2
nd
+STOP. Re-start the PC program if P1 or L1 is displayed on
the upper LCD. If the problem occurs again the next time the sys
-
tem is started, this is probably due to a configuration error.
Check whether the upper LCD on the analog display shows the drive-
on prompt (---|). In this case, automatic test routine starting has pro
-
bably not been deactivated. Inform your Service engineer.
Field bus error; please check the BNet devices.
Re-start.
End all programs. Switch the system off at the master switch box and back on
again. Re-start the PC programs.
Inform your Service engineer if the error occurs again.
No test lane created. Use the BNet tool to create a
test lane!
Test lane measurement not possible as no test lane has been defined.
Inform your Service engineer.
Program start error or
Program end error or
Test lane start error or
Test lane end error
End all programs. Switch the system off at the master switch box and back on
again. Re-start the PC programs.
Inform your Service engineer if the error occurs again.
The settings for the PC interface do not correspond
to the type of BNet converter box!
Inform your Service engineer.
No valid copy protection connector available. Check whether the copy protection connector is plugged in at the parallel port
of the PC.
It may not be possible to perform this type of measurement with the existing
copy protection connector. Consult your Service engineer.
Copy protection connector error: Timeout.
Single-user workstation:
This type of error should not really occur in this case. End all programs and
then re-start the programs.
Network:
The time difference between the storage service computer and the SDL 43xx
computer is more than 2 hours. Coordinate the times of the two computers.
End all programs on both computers and then re-start the programs.