Safety instructions
forat least one hour before being
powered on.
Q
Do not store the pump with the drive
head extended.
4.1.4
Set-up and start-up
Q
For mobile use (patient transport
within the clinic and outside the clinic)
ensure secure mounting or position-
ing of the device. Changes of position
and strong vibrations can cause minor
changes in the delivery characteristics.
Q
Ensure that the device is properly posi-
tioned and secured, and that it is level.
Q
Do not position the device above the
patient.
Q
Before powering on, check the device.
In particular, inspect the syringe
holder and claws for dirt, damage,
missing parts and to ensure that they
function correctly.
Q
Pay attention to audible and visible
alarms and the lighting up of the two
status LEDS during the self-test.
Q
When xing the device to a box rail,
donot x the device near the rail
bracket.
Q
Fully charge the battery before the rst
use without an external power supply.
4.1.5 Stacking
Q
Stack a maximum of three devices on
top of one another.
Q
Do not stack in ambulances or heli-
copters.
Q
When stacking, ensure that the device
is correctly and safely locked in. You
will hear an audible click sound when
the device is locked in.
4 Safety instructions
Q
Read the safety instructions before
using the device and observe them.
4.1 Safe handling
4.1.1 General
Q
Make sure that the introductory
training on the device is given by
a B. Braun sales representative or
another authorised person.
Q
If the device is dropped or subjected to
external forces: stop using the device
and have it tested by an authorised
service workshop.
Q
Avoid external loads on the syringe
plate sensor.
Q
Protect the device against moisture.
4.1.2
Software
Q
Consult the instructions for use fol-
lowing each software update to nd
out about the most recent changes to
the device and its accessories.
Q
Ensure that the software version on
the device corresponds to the version
these instructions for use refer to.
Q
Ensure that all devices used in a sta-
tion have the same software version
installed to avoid mistakes when using
dierently congured devices.
4.1.3
Transport and storage
Q
Do not hold the device by the drive
head during transport.
Q
Devices stored at temperatures below
the dened operating conditions range
must be kept at room temperature

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor compact plus and is the answer not in the manual?

Braun Perfusor compact plus Specifications

General IconGeneral
BrandBraun
ModelPerfusor compact plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals