Cleaning and care
10.2 Battery operation and
maintenance
The device is equipped with a modern lith-
ium-ion battery that, atthe time of delivery,
guarantees an operating time of 8hours at
5ml/h. For optimal treatment of the bat-
tery, the device is equipped with protection
against overcharge and deep depletion.
The battery is charged by the device during
mains operation. Inthe event of a power
cut or disconnection from the mains, the
pump automatically switches to battery
mode.
The battery status indicator in the display
isa trend display (low, medium, high).
10.2.1
Note for optimal battery
operation
Battery life may vary due to
Q
Ambient temperature
Q
Varying loads
Therefore, please observe the following:
Q
Under normal temperature conditions,
abattery can be fully discharged and
recharged around 300 times before its
capacity decreases to around half of
the original nominal value.
Q
When the device is in mains opera-
tion, the battery discharges slowly and
may be fully exhausted after a month
even if the device is not in operation.
Inthis case the battery does not reach
its original capacity after one charge;
ittakes several charging and discharg-
ing cycles for the battery to achieve
its original capacity.
10 Cleaning and care
Q
Device is switched o
Q
Device is unplugged from the mains
Q
Device accessories are disconnected
10.1 Cleaning
Q
No pointed objects should be used for
cleaning.
Q
Do not put excess stress on the claws
when cleaning.
Q
Clean the surface of the device with
mild soap solution.
Q
Do not spray disinfectant into the
openings in the housing.
Q
Do not use disinfectant spray on elec-
trical connections. Recommendation:
Use disinfectants manufactured by
B. Braun (e.g., Meliseptol, Melsitt10%
and Melsept SF10%) for wipe disin-
fection.
Q
Allow the device to air dry for at least
1min before operation. Donot spray
into device openings (e.g., cooling
vents, mains power plugs, interfaces).
Q
Observe all hygiene regulations.
Q
Clean accessories according to the
instructions.
Note: Substances from the groups of disin-
fectants listed below are approved, for normal
cleaning according to the manufacturer’s
instructions:
Alcohols Peroxides
QAC Active chlorine
Aldehydes Acids
Alkylamines Phenoles