Disposal
15 Start-up and
trumpet curves
15.1 Signicance in
clinical practice
Trumpet curves show the recorded maxi-
mum and minimum deviations in ow rate
compared to the delivery rate per time
interval.
In clinical practice, the trumpet curve
makes it easier for the treating doctor to
decide if the pump is suciently precise for
the administration of the desired drug.
Q
Reconcile drugs with short half lives,
inparticular, with the delivery accu-
racy in this period on the trumpet
curve.
The physiological eect of the drug can
be aected by the ow and the disposable
article.
Q
Ensure that the prescription is in line
with the start-up/trumpet curve and
the set ow rate.
13 Disposal
The device should be returned to B. Braun
for further disposal.
Q
Observe all country-specic regula-
tions when disposing of equipment
locally.
Q
Do not dispose of electrical devices
and batteries in domestic waste.
14 Safety check/service
A safety check must be performed on the
device every two years in accordance with
the checklist, with results entered into the
medical device log. The service may only be
performed by personnel who have received
training from B. Braun.