Operation
8.14 Priming the infusion line
Note: This function is not avail able in the
pump factory default. The function can
be activated by a service technician on
request.
Q
Connection to the patient removed
Q
Infusion stopped
Q
Press the Bolus key.
The “Prime infusion line” screen is
displayed.
Q
Press the up arrow key to prime the
line.
A message asking if the line is discon-
nected from the patient is displayed.
Q
Press the up arrow key to start the
priming.
The disposable item is primed with the
maximum delivery rate.
Note: After successful priming, the line can
be primed again using the up arrow key.
Q
Press the down arrow key to end the
priming.
Note: When removing a syringe if the
syringe plunger plate is not released by the
claws, the emergency release button should
be pressed. The emergency release button
is on the outside of the drive head. Itcan
be pressed using a pointed object (e.g.
ballpoint pen). Once it has been pressed
the claws can be opened by hand and the
syringe removed. Send the device to techni-
cal service.
Q
Return the syringe holder to original
position
Q
Slide the drive head towards the pump
into parking position.
8.13 Switching o the device
Q
Infusion ended
Note: The device cannot be switched o if
a disposable item is inserted. Instead it will
go into standby mode.
Ensure the drive head is in the parking posi-
tion.
Q
Press the On/O key for approx.
1.5seconds.
The device switches o.