Operation
Note: “Support for bolus-free insertion”
does not release the user from their duty of
care when changing the syringe.
Note: Always use the device with the small-
est possible syringe, provided the therapy
permits this.
Please see the notes in section 15.2
Typicalstart-up and trumpet curves.
8.3 Setting the infusion values
Q
Syringe inserted and selected
Note: Depending on the last therapy, the
pump can be set by using the delivery rate
or by using drug library.
8.3.1
Entering the delivery rate
Q
Enter the delivery rate using the arrow
keys.
Q
Start the infusion with the Start/Stop
key.
- or -
Q
Press OK to conrm the rate.
The Overview screen is displayed.
Q
Select Vol./Time and press OK
to conrm.
Q
Enter the volume or time limit and
press OK to conrm.
8 Operation
Q
Device settings congured
8.1 Switching on the device
Q
Device connected to the mains elec-
tricity or battery fully charged.
Q
Press the On/O key on the device.
The device will perform a self-test:
Note: Pay attention to audible and visible
alarms, the lighting up of the two status
LEDs and the display during the self-test.
8.2 Inserting the syringe
Q
Device switched on.
Q
Press the release lever and slide the
drive head to the right.
Q
Pull the syringe holder and turn it to
the left.
Q
Insert the syringe. Ensure that the
syringe wings have been correctly
inserted into the bracket.
Q
Pull the syringe holder and turn it to
its original position.
Q
Press the release lever and slowly slide
the drive head towards the syringe.
When the drive head reaches the
syringe plunger plate, the syringe is
automatically grasped.
The “Select syringe” message is dis-
played.
Q
Select syringe type and press OK
to conrm. Make sure that the syringe
type displayed is the same as the
inserted syringe.