30
Operation
8.8 Calculating the dose
The Dose calculation function is used to
calculate the delivery rate inml/h based on
the dose parameters entered.
Q
Syringe inserted and selected
Q
Press the Menu key.
The main menu is displayed.
Q
Select Dose calculation and press OK
to conrm.
Q
Select active substance unit and press
OK to conrm.
Q
Enter active substance quantity and
press OK to conrm.
Q
Enter volume and press OK to conrm.
The “Calculate Using” screen is dis-
played.
Calculating without patient data
The delivery rate is calculated without any
patient data being entered.
Q
Select No patient data and press OK
to conrm.
Q
Select dose unit and press OK
to conrm.
Q
Enter dose.
Note: Pressing the OK key brings up the
Overview screen.
Soft limits
Soft limits for rate/dose/bolus volume and
bolus rate can also be stored in the data-
base. These can be exceeded but the fol-
lowing message appears on the display.
The following symbols that describe the
status of the pump with regard to the soft
limits are described:
Symbol Meaning
No
symbol
Infusion is within the soft
limits
Infusion is above the upper
soft limits
Infusion is below the lower
soft limits