28
Operation
Note: Manual bolus administration is lim-
ited to a max. 10s or 10% of the syringe
content. The bolus administration is auto-
matically stopped, but it can be continued
by pressing the Bolus key again.
8.6.2
Administering a bolus with
preselected bolus volume/bolus
duration
WARNING! Danger to the patient
from an overdose. At a bolus rate of
1,200 ml/h, 1 ml is reached after 3 s.
Q
Press the OK key to stop the bolus
administration.
Q
Press the Bolus key to access the bolus
menu.
Entering the bolus volume
Q
Press the left arrow key and enter the
desired bolus volume.
Q
Press the Bolus key to start the bolus
administration.
Entering the bolus duration (optional)
Q
Press OK to conrm the entry of the
bolus volume.
Q
Select Bolus duration and press OK
to conrm.
Q
Entering the desired bolus duration.
The bolus rate is calculated.
Q
Press the Bolus key.
The bolus administration is started.
After the time has elapsed, the bolus
administration is ended and the infu-
sion continued.
8.6 Administering bolus
There are three dierent options for bolus
administration:
Q
Manual bolus
Q
Bolus with preselection of the bolus
volume
Q
Bolus with preselection of the bolus
volume and the bolus duration
Note: If the bolus administration is not
started after the Bolus key is pressed,
thedevice automatically returns to the
delivery screen for the ongoing infusion.
Note: The pressure threshold is automati-
cally increased during bolus administration.
8.6.1
Administering a manual bolus
Q
Press the Bolus key.
The “Bolus” screen is displayed.
Q
Press the Bolus key again and hold it.
Fluid is delivered as long as the key is
pressed or until the maximum duration/
dose have been reached. The delivered
bolus volume is displayed.
Q
Release the Bolus key.
The bolus administration is ended and
the infusion continued.