31
Operation
Q
Enter dose.
The rate is automatically calculated.
Note: Pressing the OK key brings up the
Overview screen.
Q
Check the plausibility of the displayed
values.
Q
Start the infusion with the Start/Stop
key.
8.9 Entering a combination
of delivery rate, volume
and time
Q
Syringe inserted and selected
Q
Press the Menu key.
The main menu is displayed.
Q
Select Rate, volume & time and press
OK to conrm.
Q
Enter two of the following parameters
and press OK to conrm:
Rate
Volume
Time
The third parameter is automatically
calculated.
If one or more parameters are entered,
changing a parameter has the following
eects on the other parameters.
Q
Rate (or dose rate) changed:
If only the volume has been
entered, the remaining time is
adjusted.
If only the time has been entered,
the remaining volume is adjusted.
If the volume and time have been
entered, the remaining time is
adjusted.
Q
Check the plausibility of the displayed
values.
Q
Start the infusion with the Start/Stop
key.
Calculate using: Weight
Q
Select Weight and press OK to conrm.
Q
Enter weight and press OK to conrm.
Q
Select dose unit and press OK
to conrm.
Q
Enter dose.
The rate is automatically calculated.
Note: Pressing the OK key brings up the
Overview screen.
Q
Check the plausibility of the displayed
values.
Q
If necessary, enter the volume or time.
Q
Start the infusion with the Start/Stop
key.
Calculate using: Body surface area
Q
Select Body surface and press OK
to conrm.
Q
Enter weight and press OK to conrm.
Q
Enter the patient’s height and then
press OK to conrm.
Q
Select dose unit and press OK
to conrm.