24
Set-up and powering on
7.3.5 Conguring service settings
Q
Select Service... and press OK
to conrm.
Q
Enter the service code and press OK
to conrm.
The “Service Menu” screen is dis-
played.
Conguring the display language
Q
Select Language and press OK
to conrm.
Q
Select the language and press OK
to conrm.
Setting the date and time
Q
Select Date and press OK to conrm.
Q
Enter the day, month and year and
press OK to conrm.
Q
Select Time and press OK to conrm.
Q
Enter the time and press OK
to conrm.
Setting the bolus rate
Q
Select Bolus rate and press OK
to conrm.
Q
Set the bolus rate and press OK
to conrm.
Alarm level Pressure value
1 0.100 bar (75 mmHg)
2 0.237 bar (178 mmHg)
3 0.375 bar (281 mmHg)
4 0.512 bar (384 mmHg)
5 0.649 bar (487 mmHg)
6 0.787 bar (590 mmHg)
7 0.925 bar (694 mmHg)
8 1.063 bar (797 mmHg)
9 1.200 bar (900 mmHg)
The set pressure level is shown with aP
(forpressure) and a number. Inaddition,
ared area shows how quickly the set
pressure alarm limit will be reached. The
“manometer” display shows the current
pressure in the system. The lower the set
pressure alarm limit level is, the larger the
red area is, the quicker this limit is reached
and a pressure alarm triggered.