EasyManua.ls Logo

Braun Perfusor compact plus - Accessories and Ordering Information

Braun Perfusor compact plus
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
54
Instructions for use for accessories
18 Instructions for use for
accessories
18.1 Interface lead 12 V
CP (8718020)
Connect the device for charging the
battery with vehicle socket
WARNING! Risk to the patient from
electric shock!
Q
Do not use the device on patients
ifthe emergency ambulance is con-
nected to the vehicle charger.
Q
Plug interface cable 12V CP into the
accessory port on the side of the
device.
Q
Plug interface cable 12V CP into the
vehicle socket.
Q
If necessary, remove the red adapter
on the vehicle socket by gently turning
it and pulling on it at the same time.
The green LED on the electronics box
shows the operating voltage.
18.2 Sta call interface lead
CP (8718030)
Connect device to the sta call system
The sta call system must comply with the
requirements of VDE 0834.
Q
Observe country-specic regulations
on sta calls.
Q
Plug the STAFF CALL interface lead
CP into the accessory port on the side
of the device or service port on the
compact
plus
station.
Q
Connect the STAFF CALL interface lead
to the sta call system.
Q
Set the sta call operating mode using
the service programme. Follow the
sta call system procedure.
Q
Check the sta call before each use of
the device.

Table of Contents

Related product manuals