Electromagnetic compatibility
Note: Distances for transmitters whose
maximum rated power is not specied in
the table above can be determined using
the equation for the relevant column, with
P being the transmitter’s maximum rated
power in watts (W) according to manufac
-
turer specications.
Note: These guidelines may not be appli
-
cable in all cases. Electromagnetic propa-
gation is aected by the absorptive and
reective qualities of the surrounding
structures, objects and people.
The compliance levels for ISM frequency
bands between 150kHz and 80MHz and
in the 80MHz to 6GHz frequency range
are designed to minimise the likelihood of
mobile/portable communications equip
-
ment causing interference if accidentally
brought into the patient area. Therefore,
theadditional factor 10/3 has been included
in the formula and used when calculating
the recommended safe distances in these
frequency ranges.
Transmitter
rated power
in W
Safe distance according to transmitter frequency m
150 kHz to 80 MHz ¹
1.2√P
80 MHz to 800 MHz
1.2√P
800 MHz to 6 GHz ¹
2.3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.27
100 12 12 23
¹ The higher frequency range applies with 80MHz and 800MHz.