40
Start-up and trumpet curves
-10
-5
0
5
10
50 ml Omnifix
Förderrate = 1 ml/h
15
-15
Epmax
Epmin
0
0,5
1,5
2,0
1
0:30 1:00 1:30 2:00
Fluss Q(t) [ml/h]
2 5 11 19
Beobachtungsfenster p x t [min]
31
Zeit [hh:mm]
0
0:30 1:00 1:30 2:00
Zeit [hh:mm]
Fluss Q(t) [ml/h]
Prozentualer Flussfehler
-10
-5
0
5
10
50 ml Omnifix
Förderrate = 5 ml/h
15
-15
2 5 11 19
Beobachtungsfenster p x t [min]
31
Prozentualer Flussfehler
Epmax
Epmin
Anlaufkurven Trompetenkurven
Flow Q(t) [ml/h]
Time [hh:mm]
Delivery rate = 1 ml/h
Flow Q(t) [ml/h]
Time [hh:mm]
Delivery rate = 5 ml/h
15.2 Typical start-up and
trumpet curves
0
0,5
1,5
2,0
1
0:30 1:00 1:30 2:00
Fluss Q(t) [ml/h]
-10
-5
0
2 5 11 19
Beobachtungsfenster p x t [min]
Prozentualer Flussfehler
31
5
10
Zeit [hh:mm]
-10
-5
0
2 5 11 19
Beobachtungsfenster p x t [min]
Prozentualer Flussfehler
31
5
10
Epmax
Epmin
Epmax
Epmin
0
0:30 1:00 1:30 2:00
Zeit [hh:mm]
Fluss Q(t) [ml/h]
Anlaufkurven Trompetenkurven
20 ml Omnifix
Förderrate = 1 ml/h
20 ml Omnifix
Förderrate = 1 ml/h
20 ml Omnifix
Förderrate = 5 ml/h
20 ml Omnifix
Förderrate = 5 ml/h
-10
-5
0
2 5 11 19
Beobachtungsfenster p x t [min]
Prozentualer Flussfehler
31
5
10
Epmax
Epmin
0:30 1:00 1:30 2:00
Zeit [hh:mm]
Fluss Q(t) [ml/h]
5 ml Omnifix
Förderrate = 1 ml/h
5 ml Omnifix
Förderrate = 1 ml/h
0
0,5
1,5
2,0
1
Start-up curves
Flow Q(t) [ml/h]
Time [hh:mm]
Delivery rate = 1 ml/h
Flow Q(t) [ml/h]
Flow Q(t) [ml/h]
Time [hh:mm]
Time [hh:mm]
Delivery rate = 1 ml/h
Delivery rate = 5 ml/h