Technical data
16 Technical data
Note: The delivery accuracy, pressure alarm and alarm reaction times apply at room tem-
perature and with water as the test material. Dierent media viscosities and temperatures
may lead to deviations.
Parameter Value
Type of device Infusion syringe pump
Product classication According to Directive 93/42 EEC:
Q
IIb
According to EN 60601-1:
Q
Protection class II
Q
For TypeCF applied parts with debrillation protection
Moisture protection IP34
Power supply
Q
100-240 V, 50-60 Hz, connection via power cable or
compact
plus
station
Q
12 V DC 12V CP interface cable
Q
10 VA typ.
Internal battery
Q
Battery life
Q
Recharging time
Lithium-ion battery
Q
Approx. 8h at 5ml/h with 50ml syringe
Q
Approx. 4h
Power consumption <20 W
Current consumption/
charging current
Q
Max. 0.6A
e
(typ. <0.1A
e
) at
100-240V, 50-60 Hz
Q
Max. 1.5A (typ. <0.5 A) at 12VDC
Sta call Max. 24V / 0.5 A / 24 VA
(VDE 0834)
EMC IEC/EN 60601-1-2 / 60601-2-24
Time of operation 100% (continuous operation)
Acoustic alarm signal
sound pressure range
Nine avail able levels:
45dB(A) to 75dB(A)