44
Technical data
16 Technical data
Note: The delivery accuracy, pressure alarm and alarm reaction times apply at room tem-
perature and with water as the test material. Dierent media viscosities and temperatures
may lead to deviations.
Parameter Value
Type of device Infusion syringe pump
Product classication According to Directive 93/42 EEC:
Q
IIb
According to EN 60601-1:
Q
Protection class II
Q
For TypeCF applied parts with debrillation protection
Moisture protection IP34
Power supply
Q
100-240 V, 50-60 Hz, connection via power cable or
compact
plus
station
Q
12 V DC 12V CP interface cable
Q
10 VA typ.
Internal battery
Q
Battery life
Q
Recharging time
Lithium-ion battery
Q
Approx. 8h at 5ml/h with 50ml syringe
Q
Approx. 4h
Power consumption <20 W
Current consumption/
charging current
Q
Max. 0.6A
e
(typ. <0.1A
e
) at
100-240V, 50-60 Hz
Q
Max. 1.5A (typ. <0.5 A) at 12VDC
Sta call Max. 24V / 0.5 A / 24 VA
(VDE 0834)
EMC IEC/EN 60601-1-2 / 60601-2-24
Time of operation 100% (continuous operation)
Acoustic alarm signal
sound pressure range
Nine avail able levels:
45dB(A) to 75dB(A)