45
Technical data
Parameter Value
Interfaces
Q
Cold connector for mains voltage
Q
Accessory port for interface cable 12V CP and sta call
Q
IrDA infrared for communication in the station and for
service
Operating conditions
Q
Temperature
Q
Relative humidity
Q
Atmospheric pressure
Q
+5 °C … +40°C (+41 °F … +104°F)
Q
30% … 90% (without condensation)
Q
0.54 … 1.06bar
Storage conditions
Q
Temperature
Q
Relative humidity
Q
Atmospheric pressure
Q
-20 °C … +55°C (-4 °F … +131°F)
Q
20% … 90% (without condensation)
Q
0.5 … 1.06bar
Weight Approx. 2.3kg
Dimensions in mm
(W x H x D)
Approx. 290 mm x 98 mm x 220 mm
(including compact
plus
stand clamp)
Safety check Every 2 years
Volume preselection 0.1ml - 9,999ml in increments of 0.01ml
Time preselection 00:01 h - 99:59 h
Delivery accuracy ±2% according to IEC/EN 60601-2-24
Occlusion alarm pressure 9 levels from 1.2 bar ± 0.2 bar.
Post occlusion bolus will be automatically reduced.
Alarm in the case of
incorrect dose
In the event of an incorrect dose of max. 0.2ml due to pump
malfunction, the pump will automatically switch o.
Max. bolus volume after
bolus reduction
≤0.2ml