EasyManua.ls Logo

Braun Perfusor compact plus - Electromagnetic Interference Emissions

Braun Perfusor compact plus
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
Electromagnetic compatibility
17.1 Electromagnetic interference emissions
Interference emission
measurements
Compliance Electromagnetic environment
guidelines
HF emissions
According to CISPR 11
Group 1 The device uses HF energy for its internal
functions only. Assuch, its HF emissions
rate is very low and it is unlikely to inter-
fere with nearby electronic equipment.
HF emissions
According to CISPR 11
Class B
The device is intended for use in all
establishments (including residential
areas and similar) directly connected to a
public power grid that also supplies build-
ings used for residential purposes.
Harmonic emissions
according to
IEC 61000-3-2
Not
applicable
Voltage uctuation/icker
emissions according to
IEC61000-3-3
Conforms

Table of Contents

Related product manuals