35
Alarms
9.2.2 Operating alarms
In the event of an operating alarm, the infusion is stopped. Anacoustic signal sounds,
thered LED ashes and a sta call is activated.
Display message Meaning
“Target volume
reached”
Preselected volume has been infused
“Disposable
syringe is empty”
No infusion solution is left in the syringe.
“Time reached” Preselected time haselapsed
“Battery empty” The battery is discharged
Q
Connect device to mains and/or have battery replaced by a
service technician
The battery alarm will sound for 3 min. Then the pump will
automatically turn o
“Pressure
too high”
There is an occlusion in the system. The set level was exceeded
Q
The pump automatically implements a bolus reduction
“KVO nished” KVO time has elapsed
“Syringe holder
open”
Syringe bracket was opened during the ongoing infusion
Q
Close syringe bracket
“Syringe not
correctly
inserted”
The wings of the syringe are not correctly inserted
Q
Insert syringe correctly, see section 8.2
“Calibrate device” Pump calibration data has changed (e.g.after an update)
Q
Recalibrate device using the service programme
“No battery
in the device”
It is not possible to use the pump without a battery
Q
Ask a service technician to insert a battery