Safety instructions
4.1.9 Enteral nutrition
The Perfusor® compact
plus
can be used for
enteral nutrition.
Q
Do not use enteral uids for the intra-
venous infusion. This would lead to a
risk of severe injury or death for the
patient.
Q
Only use disposable syringes that have
been designed and designated for
enteral nutrition.
4.2 Electrical connection
Q
Do not use the device if the plug has
visible damage.
Q
Do not use an extension cable that has
not been approved for use with device.
Q
Position the power cable so that it
does not present a trip hazard.
4.3 Safety standards
Q
The device meets all safety standards
for medical electrical equipment in
compliance with IEC/DINEN 60601-1
and IEC/DINEN 60601-2-24.
Q
It complies with the EMCthreshold
limits as specied in
IEC/DINEN60601-1-2 and
IEC/DINEN60601-2-24.
4.1.8 Accessories and consumables
Q
Change the disposables according to
your local infection control policy.
Q
Only use pressure-tested disposable
items (min.2bar/1,500mmHg).
Q
Only use the device with accessories
and consumables that have been
approved for use with the device.
Q
Ensure·adequate·protection·against·f
ree-ow before changing disposable
items.
Q
Always use the device with the small-
est possible syringe, provided the
therapy permits this.
Q
See the corresponding manufacturer
information for possible incompatibili-
ties between the device and medicinal
products.
Note: The use of untested or incompatible
disposable items can aect the technical
data.
Q
Use only Luer lock feed systems and
syringes as well as compatible device,
accessory, wear part and disposable
item combinations.
Q
Connected electrical components must
comply with IEC/EN specications
(e.g., IEC/EN 60950 for data process-
ing equipment). Anyone who connects
additional devices is considered a
system congurer, and is therefore
responsible for compliance with sys-
tem standard IEC/DINEN60601-1-1.
Q
If more than one appliance/infusion
line is connected, mutual interference
cannot be ruled out
.