Electromagnetic compatibility
17.2 Electromagnetic immunity
The device is designed to be used in the electromagnetic environment described below. The
device users and customers should ensure that it is being operated in such an environment.
Immunity tests Test level
EN 60601-1-2
EN 60601-2-24
Compliance
level
Electromagnetic
environment guidelines
Electrostatic
discharge (ESD)
according to
IEC 60601-4-2
Contact discharge
EN 60601-1-2:
±6 kV
IEC 60601-2-24:
±8 kV
±6 KV
without
interference
±8 KV outage
with alarm
permitted
Floors should be wood, con-
crete, orceramic tile. Ifthe
oor covering is made of a
synthetic material, relative
air humidity needs to be at
least 30%.
Air discharge
EN 60601-1-2:
±8 kV
IEC 60601-2-24:
±15 kV
±8KV without
interference
±15KV outage
with alarm
permitted
Electrical fast
transient/ bursts
according to
IEC60601-4-4
for power supply
lines ±2 kV
±2 kV The supply voltage quality
should be the same as that
of a typical commercial or
hospital environment.
For input and output
lines ±1 kV
±1 kV
Surges according
to IEC 61000-4-5
±1 kV outer conduc-
tor - outer conduc-
tor voltage
±1 kV The supply voltage quality
should be the same as that
of a typical commercial or
hospital environment.
±2 kV voltage
Outer conductor -
ground
±2 kV