46
Technical data
Parameter Value
KVO rate
Q
Rate: ≥ 10ml/h: KVO rate 3ml/h
Q
Rate: < 10ml/h: KVO rate 1ml/h
Q
Rate: < 1ml/h: KVO rate = rate set using the service
program (factory default rate 0.1ml/h) or current rate if
this is lower.
History protocol
Q
1,000 history entries
The oldest entries are overwritten if necessary.
Q
100 events for system diagnosis
The history is retained when the device is switched o or
the battery removed.
Delivery rates
Continuous delivery rates/bolus rates according to the syringe size used:
Syringe size [ml] Continuous
delivery rate [ml/h]
Bolus rate [ml/h] Preset bolus rate
[ml/h]
50/60 0.01 to 200
Or alternatively:
0.01 to 999.9
1 to 1800 800
30/35 0.01 to 100 1 to 1200 600
20 0.01 to 100 1 to 800 400
10/12 0.01 to 50 1 to 500 200
5/6 0.01 to 50 1 to 300 150
2/3 0.01 to 25 1 to 150 80
Note: The delivery rate can be set in steps of 0.01ml.
Note: The preset bolus rate can be changed via the service menu or once via the combination
of bolus volume and bolus time.
Delivery rate accuracy in bolus administration is generally ± 2%. The accuracy can vary
when administering low bolus volumes.