51
Electromagnetic compatibility
Immunity tests Test level
EN 60601-1-2
EN 60601-2-24
Compliance
level
Electromagnetic
environment guidelines
Radiated HF
interference
according to IEC
61000-4-3
10 V/m
80MHz to
2.5 GHz
[E1] 10 V/m
80MHz to
6GHz und
500MHz to 3
GHz
The eld strength should be
lower than 10V/m
d = 12/E1 √P ²
80MHz to 800MHz
d = 23/E1 √P ²
800MHz to 6 GHz
Field strengths from
stationary RF transmit-
ters should be below the
compliance level for all
frequencies, based on an
on-site test.
Interference is possible in
the vicinity of equipment
that has the following
symbol.
¹ UT is the AC mains voltage prior to test level application
² With P as the maximum rated power of the transmitter in watts (W) according to the trans-
mitter manufacturer specications and as the recommended safe distance in metres (m).

Table of Contents