Brookeld Engineering Labs., Inc. Page 47 Manual No. M13-167
The GlobalAlarm is deactivated by choosing None in the drop down list of
availablemeasurementparameters.YoumustpressOKtoconrmyourGlobal
Alarmchoice.
Spindle List:TheselectionofSpindleinCongureViscosityTestcanbedonebyusingeithera
numberpadorascrolllist.(Thischoiceisselectedthroughthenavigationbutton
intheSetSpindlescreen.)ThecontentsofthescrolllistaredenedintheSpindle
Listmenu.Eachspindlecanbeindividuallyselectedordeselectedthroughtheuse
ofacheckbox.Spindles,shownwithboththeirnameandtheirnumber,arelisted
numericallyaccordingtotheirspindleentrynumber.
Therestrictedspindlewillnotappearonthescrolllistofavailablespindleswhen
choosingSpindleinCongureViscosityTest.Iftheuserselectstoenterthespindle
by number, then the choice of a restricted spindle will present an Invalid Input
messageandpreventtheselectionofthespindle(seeFigureIV-8).
List Button
Figure IV-8
TIP: Toensurethatthecorrectspindlecodeisalwaysentered,createatestthat
issavedtotheDV2Tmemory.
TIP:TheselectionofspindlecanberestrictedbyUserLogIn.Usethisrestriction
combinedwithasavedtestcongurationtoensurethatthecorrectspindle
isalwaysusedforcriticaltests.SeeSectionIV.4.2fordetailsofUsersand
Access.
ASpecialSpindlecanbecreatedwhentheuserhasdevelopedanewspindlegeometry
orwhenusingastandardBrookeldspindleinanonstandardmeasurementcontainer.
TheSpecialSpindle,oncecreated,willbeavailabletousersthroughtheSpindle
selectioninCongureViscosityTest.
ASpecialSpindleiscreatedbyselectingtheSpecialSpindlecommandbuttoninthe
EditSpindleListmenu.ASpecialSpindlerequiresacode,aname,andanSMC
value.OptionalvaluesincludeSRCandYMC.