Brookeld Engineering Labs., Inc. Page 48 Manual No. M13-167
Code Thecodeisthenumericalvalueusedtoselectthespindle.Thisvalue
mustbeuniqueandcannotbethesameasastandardBrookeld
spindle.TheacceptablerangeforCodeis100-199.
Name TheNameisauniquevaluetodescribethespindle.TheNamewill
bedisplayedintheSpindleeld.TheNamecanbecomprisedof
alphaandnumericcharacters.
SMC TheSMC(SpindleMultiplierConstant)isusedtoconvertthemeasured
torquetoviscosity.TheacceptablerangeofSMCis0.001–500.
The SMC value can be determined through the use of the Range
equationfoundinappendixD.Someexperimentationwiththespindle
andacalibratedviscositystandardisrequired.
Range(cP)=TK*SMC*10,000/RPM
TheTorqueConstant(TK)canbefoundinTableD-2inAppendixD.
SRC TheSRC(ShearRateConstant)isusedtoconvertthesetRPMto
shearrateandcalculateShearStress.TheacceptablerangeofSRC
is 0.001 – 500.
TheSRCvaluecanbedeterminediftheshearratecharacteristicsof
thespindlegeometryareknown.ConsulttheBrookeldpublication,
“MoreSolutionstoStickyProblems”,foradiscussionaboutshear
rateandsomerecommendedequations.
ShearRate(1/s)=SRC*RPM
YMC TheYMC(YieldMultiplierConstant)isusedtoconvertthemeasured
torquetoyieldstress.
This feature is only available with the DV3T Rheometer. Consult
the DV3T operating instructions for more information.
Speed List: TheselectionofSpeedinCongureViscosityTestcanbedonebyusingeithera
numberpadorascrolllist.(Thischoiceisselectedthroughthenavigationbuttonin
theSetSpindlescreen.)Whenscrolllistisselected,thespeedentryisrestrictedtothe
classicBrookeldspeedsetsfortheLVandRV/HA/HBviscometers.Thecontents
ofthescrolllistaredenedintheSpeedListmenu.Eachspeedcanbeindividually
selectedordeselectedthroughtheuseofacheckbox.Selectordeselectallspeeds
byusingtheSelectAllcheckbox.SelectordeselectallLV(orRV/HA/HB)speeds
byusingtheSelectAllLVcheckbox.TheclassicBrookeldspeedsetsaredened
below.
LV: 0.3,0.6,1.5,3,6,12,30,60
RV/HA/HB: 0.5,1.0,2.0,2.5,4,5,10,20,50,100
Additionally,anewspeedcanbeaddedtothelistbyselectingtheNewSpeedcommand
keyatthebottomofthedisplay.TheNewSpeedmustbeselectedwithintherange
ofavailablespeedsontheDV2T:0.0–200RPM.
Speedentrythroughthenumberpadisneverrestricted.
TIP: Toensurethatthecorrectspeedisalwaysentered,createatestthatissaved
totheDV2Tmemory.
TIP: TheselectionofspeedcanberestrictedbyUserLogIn.Usethisrestriction
combinedwithasavedtestcongurationtoensurethatthecorrectspeed
isalwaysusedforcriticaltests.SeeSectionIV.4.2fordetailsofUsersand
Access.