Troubleshooting
56 Z33081_00_001
Problem Cause Remedy
Lowcoolantlevel. Refillcoolantreservoir(see
sectionCryocooler Coolant
Circuit [}58]).
Radiatordirty. Cleantheradiatorwitha
vacuumcleaner(see
sectionCleaning Instructions
[}58]).
Incorrectfanspeed. ContactBrukerservice.
Defectivewaterpump. ContactBrukerservice.
Coolanttemperature
sensorfailure.
ContactBrukerservice.
Whenready,cannotdeliver
parahydrogenincontinuous
flowmode.
Conversionchamber
blockage.
Executeawarmup(section
Warm Up [}53]),thena
purge(Purge Function [}48]
).
Outletvalvemalfunction. ContactBrukerservice.
Batchdeliverycontainer
evacuationfails.
Leakincontainer
connections.
Check/remakeall
connectionstothe
container.
Timeoutwhilefilling
container.
Conversionchamberis
blocked.
Executeawarmup(section
Warm Up [}53]),thena
purge(Purge Function [}48]
).
Repeatedconversion
chamberblockages.
H
2
gaspurityproblem. UsehighpurityH
2
gas:
99.999%orbetter.
Purgenotexecutedpriorto
cooldown.
Executeawarmup(section
Warm Up [}53]),thena
purge(Purge Function [}48]
).
Purifiercartridgesaturated. ContactBrukerservice.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker BPHG 90 and is the answer not in the manual?

Bruker BPHG 90 Specifications

General IconGeneral
CategoryLaboratory Equipment
TypeHigh Pressure Homogenizer
Magnetic Field StrengthNot Applicable
Proton FrequencyNot Applicable
Flow RateData unavailable
Temperature RangeData unavailable

Related product manuals