58
Operating and commissioning
Type 8228
OK
Save or not the calibration result by
choosing "Yes" or "No".
The message "Error: out of range" signals that the cell
constant is out of the authorized range (< 0,8 or > 12);
This may be due to:
• either a mistake made when entering the conductivity,
• or when there are air bubbles in the hole of the conduc-
tivity sensor
• or when the minimum distance of 4 cm between the
conductivity sensor and the sides of the beaker is not
observed.
BACK
Calib Sensor CalibrationProbe
→ Immerse the clean conductivity sensor in the solution with a known
conductivity. The device alternately displays:
• the measured conductivity of the solution
• the measured temperature of the solution
→ Enter the conductivity, at
the fluid temperature, of
the reference solution used
(marked on the bottle or
measured using a reference
instrument).
→ Change the unit if necessary.
Calibration
5.023 µS/cm
5.000 µS/cm
Cal. Result
+1.00000
The device displays the calibration
result.
OK
Cal. Result
Save: Yes/No
"Error: out of
range"
Probe
English