Maintenance and troubleshooting
Type 8228
Red
LED
Current
output
Transistor
output
Icon Message dis-
played in the
Info menu
Possible cause Recommended action
ON 22 mA depending on
thresholds
ERR
+
"TR EE Fact
Read"
Parameter reading error. Switch the device off
and on again.
If the error persists,
set the device back
to the default settings
(chap. 9.11.4).
If the problem per-
sists, return the device
to Bürkert.
"TR EE User
Read"
ON 22 mA depending on
thresholds
ERR
+
"TR COM
Measure"
The acquisition/con-
version module of the
process values is faulty.
The process is stopped.
Switch the device off
and on again.
If the problem per-
sists, return the device
to Bürkert.
ON 22 mA depending on
thresholds
ERR
+
"TR EE
UserWrite"
Parameter saving error. Switch the device off
and on again.
Save the settings
again.
If the error persists,
set the device back
to the default settings
(chap. 9.11.4).
If the problem per-
sists, return the device
to Bürkert.
OFF 4-20 mA depending on
thresholds
+
"S RTC Reinit" The date and time are
lost because the device
has not been powered
up for at least 3 days.
Set the date and
time again (see
chap. 9.11.2).
Feed the transmitter
for at least 10 minutes
so that the date and
time are battery fed
for 3 days.