Proper sizing is critical for effective therapy. Use the nasal
prong and nasal mask sizing guide to select the correct
interface size. The nasal prongs and masks are color-coded
for quick size identification and match the sizing guide
colors. The size of the interface is also located at the base
of the nasal prongs and mask. The base is keyed to ensure
the proper placement of the interface onto the generator.
1. Align the indentation at the base of the nasal prong
or mask over the pressure line. Press it evenly onto the
generator, and check it for a tight fit.
2. Set the flow rate to deliver the prescribed pressure.
3. Occlude the prongs or mask, and verify that the CPAP
level is reached. If the desired CPAP level is not achieved,
check for leaks.
4. Remove the support cradle from the generator assembly.
Tip: Routine care: Choosing nCPAP over intubation can
provide a gentler therapy option. It also introduces some
delivery challenges as the delicate skin of premature
infants is extremely susceptible to damage from excessive
pressure or abrasion. The infants nasal area must be
routinely inspected for skin irritation and breakdown;
if indicated, discontinue therapy. Periodically alternate
between the nasal prongs and mask to help reduce the
incidents of skin breakdown.
Routinely inspect/visualize the area under the fixation
device for skin irritation. Remove and/or adjust the fixation
device as needed to ensure proper placement.
Generator assembly preparation
Following the facilitys protocol, set up the Infant Flow SiPAP
or nCPAP driver and attach a breathing circuit. The heated,
single-limb infant breathing circuit must be compatible
with the humidifier. Remove the Infant Flow LP generator
from the package, and ensure the corrugated tubing is fully
extended. Firmly attach the pressure line of the breathing
circuit to the generator assembly pressure line. Next, attach
the breathing circuit supply line to the drive line of the
generator assembly. Ensure the fittings are tight. Attach
the nasal prongs or nasal mask to the generator.
A second clinician may be needed to assist during
the application process.