Bonnet application (alternative method 2)
A second method for applying the bonnet is attaching the
generator assembly to the bonnet prior to application. All
three methods discussed are acceptable ways to apply the
Attach the generator assembly to the bonnet
Another option for applying the bonnet is to attach
the generator assembly, interface and support cradle
prior to application.
1. Prepare the generator assembly as indicated in the
previous section. Attach the desired interface and
connect the gray straps to the quick secure tabs.
2. Center the support cradle on the application pad.
3. Place the drive and pressure lines in the support cradle
with the exhaust tube resting on top. Wrap the securing
strap around the exhaust tube, and attach it to the hook
material on the side of the support cradle. Keep a little
slack in the securing strap to freely move the generator
when ready to position it.
4. Thread the side straps onto the bonnet to fasten the
generator assembly to the bonnet.
5. Pull the bonnet and generator assembly down over
the infants head. Make any necessary adjustments to
properly position the generator and interface over the
infant’s nasal area.
6. Insert the nasal prongs or position the mask. Adjust the
tension of the side straps.
7. Complete a final inspection of the set-up. Confirm the
infant’s ears are covered and the bonnet’s edge is at the
brow line. Check that the interface is properly positioned
and a good seal is achieved.
Note: The preparation of the bonnet and generator
assembly can be completed prior to application. The
actual process of applying it to the infant may be a little
cumbersome. It takes practice to keep the generator
assembly from brushing against the infant’s face while
pulling the bonnet down over the head. Ask another
caregiver to assist during the application process.
bonnet and generator assembly. Clinicians should use
the method that they are most comfortable with and best fits
the situation.