CS 7200 User and Installation Guide (6M0940)_Ed04 45
1 Select the X-ray exposure time according to the region of interest and the patient type. Follow the user
instructions of your X-ray generator. The following tables provide guidelines for exposure times for an
X-ray generator at 70 kV and 7mA. Add your values for the exposure time in seconds in the column on
the right.
Table 3 Exposure times
Parameter DC-emitter, 7mA, tube length 20 cm DC-emitter, 7mA, tube length 30 cm
recommended exposure time (s) recommended exposure time (s)
Upper jaw 60kV 70kV 60kV 70kV
incisors 0.1 0.08 0.2 0.16
pre-molars 0.125 0.1 0.25 0.2
molars 0.16 0.125 0.32 0.25
Lower jaw 60kV 70kV 60kV 70kV
incisors 0.1 0.08 0.2 0.16
pre-molars 0.125 0.1 0.25 0.2
molars 0.125 0.1 0.25 0.2
Bitewing 0.16 0.125 0.32 0.25
Important: Make sure the active side of the imaging plate is facing the
X-ray tube.
Important: These are suggested exposure times and need to be adjusted
for your specific X-ray generator.
Tip: If images are too dark, reduce the exposure time. If images are too white
or look grainy, increase the exposure time.