137
User’s Manual
CellaVision
®
DM1200
(Cont'd)
Accuracy results for CellaVision DM1200 with the body fluid application:
Short-term imprecision results for CellaVision DM1200 with the body fluid
application:
(Cont'd)
Specimen Cell class Slope Intercept (%)
R
2
# of samples
CSF
Neutrophils 1.0005 0. 16 0.9919 62
Lymphocytes 0.9703 0. 27 0.9897 62
Eosinophils 0.7573 0. 05 0.7893 62
Macrophages 0.9944 -0. 44 0.9821 62
Other cells 0.9843 1.15 0.9435 62
Serous fluid
Neutrophils 0.9967 0. 63 0.9896 151
Lymphocytes 0.9879 -0. 28 0.9813 151
Eosinophils 1.1326 -0. 002 0.9763 151
Macrophages 1.0099 -0. 37 0.9809 151
Other cells 0.9432 0. 06 0.9261 151
Synovial fluid
Neutrophils 1.0007 0. 15 0.9959 47
Lymphocytes 0.9891 1.10 0.9786 47
Eosinophils 0.7701 0.04 0.9262 47
Macrophages 0.9551 0.06 0.9586 47
Other cells 0.5027 0.19 0.1735 47
Specimen Cell class SD (%)
CSF
Neutrophils 3.1
Lymphocytes 3.9
Eosinophils 0.6
Macrophages 3.0
Other cells 2.8
Serous fluid
Neutrophils 3.2
Lymphocytes 6.3
Eosinophils 0.9
Macrophages 6.7
Other cells 1.1
Synovial fluid
Neutrophils 3.3
Lymphocytes 5.3
Eosinophils 0.1
Macrophages 5.6
Other cells 1.0