EasyManuals Logo

Cessna 177 SERIES Service Manual

Cessna 177 SERIES
430 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #222 background imageLoading...
Page #222 background image
alternate
static
source
valve
(if
installed).
piece
of
tape
over
small
hole
in
lower
aft
end
of
pitot
c.
Attach
a
source
of
suction
to
static
pressure
tube,
fasten
a
piece
of
rubber
or
plastic
tubing
over
source
opening.
Figure
15-3
shows
one
method
of
pitot
tube,
close opposite
end
of
tubing
and
slowly
obtaining
suction.
roll
up
tube
until
airspeed
indicator
registers
in
d.
Slowly
apply
suction
until
altimeter
indicates
a
cruise
range.
Secure
tube and
after
a
few
minutes
1000-foot
increase
in
altitude,
recheck
airspeed
indicator.
Any
leakage
will
have
reduced
the
pressure
in
system,
resulting
in
a lower
CAUTION
airspeed
indication.
Slowly
unroll
tubing
before
removing
it,
so
pressure
is
reduced
gradually.
Other-
When
applying
or
releasing
suction,
do
not
wise
instrument
may
be
damaged.
If
test
reveals
a
exceed
range
of
vertical
speed
indicator
or
leak
in
system,
check
all
connections
for
tightness.
airspeed
indicator.
e.
Cut off
suction
source
to
maintain
a
"closed"
15-15.
BLOWING
OUT
LINES.
Although
pitot
sys-
system
for
one
minute.
Leakage
shall
not
exceed
tern
is
designed
to
drain
down
to
pitot tube
opening,
100
feet
of
altitude
loss
as
indicated
on
altimeter.
condensation
may
collect
at
other points
in
system
f.
If
leakage
rate
is
within
tolerance,
slowly
re-
and
produce
a
partial
obstruction.
To
clear
line,
lease
suction
source.
disconnect
at
airspeed
indicator.
Using
low
pres-
sure
air,
blow
from
indicator
end
of
line
toward
NOTE
pitot
tube.
If
leakage
rate
exceeds
maximum
allowable,
iCAUTIONI
first
tighten
all
connections,
then
repeat
leakage
test.
If
leakage
rate
still
exceeds
Never
blow
through
pitot
or
static
lines
maximum
allowable,
use
following
procedure.
toward
instruments.
g.
Disconnect
static
pressure
lines from
airspeed
Like
pitot
lines,
static
pressure
lines must
be
kept
indicator
and
vertical
speed
indicator.
Use
suitable
clear
and
connections
tight. Static
source
sumps
fittings
to
connect
lines together
so
altimeter
is
the
collect
moisture
and
keep
system
clear.
However,
only
instrument
still
connected
into
static
pressure
when
necessary,
disconnect
static
line
at
first
in-
system.
strument
to
which
it
is
connected, then
blow
line
h.
Repeat
leakage
test
to
check
whether
static pres-
clear
with
low-pressure
air.
Check
all
static
pres-
sure
system
or
the
bypassed
instruments
are
cause
of
sure
line connections
for tightness.
If
hose
or
hose
leakage.
If
instruments
are
at fault,
they
must
be
connections
are
used,
check
for
general
condition
and
repaired
by
an
"appropriately
rated
repair
station"
clamps
for
security.
Replace
hose
which
have
crack-
or
replaced.
If
static
pressure
system
is
at
fault,
ed,
hardened
or
show
other signs
of
deterioration.
use
following
procedure
to
locate
leakage.
i.
Attach
a
source
of
positive
pressure
to
static
15-16.
REMOVAL
AND
INSTALLATION
OF
COM-
source
opening.
Figure
15-3
shows
one
method
of
PONENTS.
(Refer
to
figure
15-2.)
To
remove
pitot
obtaining
positive
pressure.
mast
remove
four
mounting
screws
on
side
of
connec-
tor
(11)
and
pull
mast
out
of
connector
far
enough
to
jCiAUTIONS
disconnect
pitot
line
(13).
Electrical
connections
to
heater
assembly
(if
installed)
may
be
disconnected
Do
not
apply
positive
pressure
with
airspeed
through
wing
access
opening
just inboard
of
mast.
indicator
or
vertical
speed
indicator
connect-
Pitot
and
static
lines
are
removed
in
the
usual
manner,
ed
to
static
pressure
system.
after
removing
wing
access
plates,
lower
wing
fairing
strip
and
upholstery
as
required.
Installation
of
j.
Slowly
apply
positive
pressure
until
altimeter
tubing
will
be
simpler
if
a
guide
wire
is
drawn
in
as
indicates
a
500-foot
decrease
in
altitude
and main-
tubing
is
removed
from
wing.
The
tubing
may
be
re-
tain
this
altimeter
indication
while
checking
for
leaks.
moved
intact
by
drawing
it
out
through
cabin
and
right
Coat
line
connections
and
static
source
flange
with
door. Tighten
connections
firmly
but
avoid
over-
solution
of
mild
soap
and
water,
watching
for
bubbles
tightening
and
distorting
fittings.
If
twisting
of
plas-
to
locate
leaks.
tic
tubing
is
encountered
when
tightening
fittings,
k.
Tighten
leaking
connections.
Repair
or
replace
VV-P-236
(USP
Petrolatum)
may
be
applied
sparingly
parts
found
defective.
between
tubing
and
fittings.
1.
Reconnect
airspeed
and
vertical
speed
indicators
into
static
pressure
system
and
repeat
leakage
test
per
steps
"c"
thru
"f".
15-14.
PITOT
SYSTEM
INSPECTION
AND
LEAKAGE
TEST.
To
check
pitot
system
for
leaks,
place
a
15-6
Change
5

Table of Contents

Other manuals for Cessna 177 SERIES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cessna 177 SERIES and is the answer not in the manual?

Cessna 177 SERIES Specifications

General IconGeneral
ManufacturerCessna
Engine TypePiston
Wingspan35 ft 6 in (10.82 m)
Seating Capacity4
Number Built4, 295
EngineLycoming O-360
Horsepower180 hp
Maximum Takeoff Weight2, 500 lb (1, 134 kg)

Related product manuals