EasyManua.ls Logo

Chino DB650 - Input Correction; Limiter

Default Icon
150 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
[DB600] General Instruction Manual MODE 5
- 90 -
5-7-4
Input correction
Sets the sensor correction and digital filter.
Sensor correction
Adds a correction value to measured value (PV) and uses the result for the instrument display or control.
This function provides a fine adjustment for the case a variation of measured value is observed between
the instrument and other instruments.
Digital filter
Reduces variation in measured value (PV) caused by noise, etc. to make it smooth. A first-order lag
operation is performed on measured value to set the time (time constant) to be approximately 63%. The
larger the time constant, the more effective the noise removal performance; however, input response
becomes slow if the time constant is too large.
5-7-5
SV limiter
SV limiter is used to limit the SV operation range (lower/upper limit). It is functions in program operation
and constant value operation. SV limiter does not function in remote operation.
Setting/changing of SV in MODE0, MODE2 or MODE3 should be done within the setting range of SV
limiter.
Time
Time constant (0.1 to 99.9sec, 0.0 = OFF)

Table of Contents

Related product manuals