5.7.2 Focus Num
In B mode, 4 focus points can be selected simultaneously, and the number controlled by the depth,
SRA and Compound.
Move the cursor to [Focus Num] selection to adjust, and the range is 1~4. When Compound or
SRA is on, Focus Num can be adjusted to 2 (max).
5.7.3 Persistence
In real status, adjust the contrast and resolution.
In real status, press the first line button of [Persistence] to adjust. The range is 0~7.
5.7.4 Line Density
Scan Line Density function is only valid for the image in B mode, 2B mode, B/M mode or 4B mode
image. The line density has two types: high density and low density. High density means better
image quality while low density image has higher frame rate.
To do the adjustment, please select the menu item [Density] and press [MENU] to adjust the line
density.
5.7.5 Smooth
Smoothness function is used for restraining the image noise and performing axial smooth
processing to make the image smoother.
Move the cursor to [Smooth] selection to adjust, and the range is 0~7.
5.7.6 Edge Enhance
Edge enhancement is used for enhancing the image outline. In this way the user can view the
tissue structure more clearly.
Move the cursor to [Edge] selection to adjust, and the range is 0~7.
5.7.7 Acoustic Power
Acoustic power means the acoustic power transmitting from the probe.
At the real-time status, Move the cursor to [Acoustic power] selection to adjust, and the range is
0~15.
5.7.8 Utility
This function includes post processing, slide show and other items. Press [MENU] button, then
choose Utility selection, it will appear Utility options.